Disposable Administration Set Illustrations and Features
CAUTION: USE ONLY CURLIN MEDICAL ADMINISTRATION SETS
WITH THIS PUMP
Figure 1.3 Samples of Curlin Medical Administration Sets
These drawings are
NOT to scale.
Note: An important feature of the Curlin Medical administration set is the unique
design of the set-based “Integral Flow-Stop.” When the set is new, the “Integral
Flow-Stop” has a breakaway tab that keeps the “Integral Flow-Stop” open and allows
the tubing to be gravity primed. When you are ready to load the administration set
into the pump, remove this tab, and the “Integral Flow-Stop” will automatically clamp
the tubing. The Curlin Medical “Integral Flow-Stop” prevents inadvertent flow of
medication to the patient whenever the door of the pump is opened because the
“Integral Flow-Stop” automatically reacts to clamp the tubing. The “Integral Flow-
Stop” can, however, be opened after the tab is broken away by intentionally squeezing
down on the movable spring action section of the “Integral Flow-Stop.” See Figure 1.6
for an illustration of how to intentionally open the “Integral Flow-Stop.” For a
complete listing of all available administration sets, see the Curlin Medical Product
Catalog.
Integral Flow-
Stop with
breakaway tab
T
ubing Guide
Bag Spike
Slide Clamp
Male Luer
Lock
CURLIN
MEDICAL
Integral Flow-
Stop with
breakaway tab
T
ubing Guide
Medication
Reservoir
Slide Clamp
Male Luer
Lock