MIQ/CR3; DIQ/CR3 Indexes
8 - 1
ba76032e01 01/2012
8Indexes
8.1 Explanation of the messages
In this chapter you will find a list with all the message codes and
corresponding message texts that may occur in the log book of the
IQ S
ENSOR NET system for the MIQ/CR3 output module.
Note
Information about
Contents and structure of the log book and
Structure of the message code
can be found in the L
OG BOOK chapter of the IQ SENSOR NET system
operating manual.
Note
All message codes of the output module MIQ/CR3 or DIQ/CR3 end
with the number "411".
8.1.1 Error messages
8.1.2 Info messages
The MIQ/CR3 or DIQ/CR3 output module does not send any info
messages.
Message code Message text
EA4411 Current output range undercut
* Check process
* Check settings and, if necessary, change them
EA5411 Current output range exceeded
* Check process
* Check settings and, if necessary, change them
EI3411 Burden resistor too large (> 500 Ohm) or current loop interrupted
* Check burden, terminal connections and connection lines